User Contributed Met/Cal Procedure ============================================================================== INSTRUMENT: Electro Scientific Industries 1701A Ohmmeter. DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: 0.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 13 NUMBER OF LINES: 64 CONFIGURATION: Fluke 5450A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J 1.002 DISP Let the unit warm up for at least 15 minutes if you 1.002 DISP have not done so already. 1.003 SET RATE/FAST: In 1.003 SET CYCLE HOLD: Out 1.004 DISP Set RANGE TO 20 m[24]. 1.005 DISP Short the Test Leads. 1.006 MEMI Enter the reading. 1.007 MEMC 0Z 0.011U 2.001 DISP Set the RANGE to 200 m[24]. 2.002 DISP Connect the leads to one of the Cal. Lab's 0.1[24] 2.002 DISP standard resistors. 2.003 MEMI Enter the meter reading. 2.004 MEME 2.005 MEMI Enter the resistor's value in milliohms. 2.006 MEMC mZ 0.02U 0.02% 3.001 DISP Set the RANGE to 2[24]. 3.002 5450 1Z .0001U .02% 4W 4.001 5450 1.9Z .0001U .02% 4W 5.001 DISP Set the RANGE to 20[24]. 5.002 5450 10Z .001U .02% 4W 6.001 5450 19Z .001U .02% 4W 6.002 5450 19Z S 4W 6.003 DISP Adjust R CAL through the back panel to make the meter 6.003 DISP read the same as the stimulus. 6.004 JMP 7.002 7.001 JMP 7.002 5450 19Z .001U .02% 4W 8.001 5450 10Z .001U .02% 4W 9.001 DISP Set the RANGE to 2[24]. 9.002 5450 1.9Z .0001U .02% 4W 10.001 5450 1Z .0001U .02% 4W 11.001 DISP Set the RANGE to 200 m[24]. 11.002 DISP Connect to the SAME standard resistor as before. 11.003 MEMI Enter the meter reading. 11.004 MEME 11.005 MEMI Enter the resistor's value in milliohms. 11.006 MEMC mZ 0.02U 0.02% 12.001 DISP Set the RANGE to 20 m[24]. 12.002 DISP Short the leads. 12.003 MEMI Enter the reading. 12.004 MEMC 0Z 0.011U 13.001 DISP Now, wasn't this a weird piece of shi---(gear)? END